Team:NYMU-Taipei/Project/Time Regulation/Results of Reloxilator

From 2008.igem.org

Revision as of 01:58, 30 October 2008 by Blackrabbit (Talk | contribs)

Only the Oscillator, Tuner, and Reporter was successfully ligated together, so we performed a few Reporting Assays on it.

Contents

Reporting Assay 1

This was done having little knowledge of what we were actually trying to look for. We just needed some data to get a feel for what kind of information we were looking for, and we also used this experiment as an indicator to how much time the experiment would take and what we could improve upon next time (e.g. how to take the samples from the tubes and put it back in the incubator in the shortest amount of time in order not to affect the experiment too much).

Method

Diluted all samples by a 1:100 ratio and let them grow. Measured the OD and fluorescence every hour.

Results

  1. Found that doing technical replicates actually makes a lot of difference if you didn't pipette properly.
  2. If starting with and OD600 of around 0.0325, the fast growth period where we see the most activity lies between 2 to 6 hours. From the next experiment on, we decided to dilute all samples to start off with an OD around 0.0325.
  3. Other than that, at first we had thought that the fluorescence values were too close to the blank (LB medium), and some even lower than it, so we did not analyse the data. It wasn't until after the third reporting assay we noticed that the measured points did indeed have a sinusoidal shape.

Reporting Assay 2

Reporting Assay 3

Reporting Assay 4

Started doing reporting assay of the Oscillator+Tuner+Synchronizer+Reporter construct, until we found out halfway that the construct had actually failed (we were using the time between measurement assay points to validate the length of the part from plasmid extraction->digestion->gel electrophoresis.

Reporting Assay 5

  • Aim
    • purely wanted to see the relationship that different amounts of IPTG has upon the fluorescence of the Oscillator+Tuner+Reporter construct. So OD600 was not measured except at the start.
  • IPTG concentrations: 0,50,100,200,300,500,1000,2000,4000 uM.
  • Measurement: every 5 minutes at the wavelengths (ex/em):
    • 488/533
    • 488/525
    • 475/515
    • 470/530

Results

There was no significant difference between the wavelengths used. The shape looks the same to the human eye.