Standard Curve in ddH2O
7/23, 2008 PM 4:04
- 100, 200, 300, 400, 500 uM pi
- reaction time: 10 mins
- 5N H2SO4
|
7/30, 2008 PM 8:00
- 500, 1000, 1500, 2000, 2500, 3000 uM pi
- reaction time: 11 mins (the timing of measure is also important)
- 5N H2SO4
- Because the high [pi] is out of the linear range, we have to dilute the samples into linear region
- standard samples are diluted from 10,000 uM stock
- The dilution procedure are:
- extract 30 uL from 10,000 uM stock + 70 uL ddH2O, and add 900 uL ddH2O to 1 C.C. (3000uM, 10X dilution)
- extract 25 uL from 10,000 uM stock + 75 uL ddH2O, and add 900 uL ddH2O to 1 C.C. (2500uM, 10X dilution)
- extract 20 uL from 10,000 uM stock + 80 uL ddH2O, and add 900 uL ddH2O to 1 C.C. (2000uM, 10X dilution)
- extract 15 uL from 10,000 uM stock + 85 uL ddH2O, and add 900 uL ddH2O to 1 C.C. (1500uM, 10X dilution)
- extract 10 uL from 10,000 uM stock + 90 uL ddH2O, and add 900 uL ddH2O to 1 C.C. (1000uM, 10X dilution)
- extract 05 uL from 10,000 uM stock + 95 uL ddH2O, and add 900 uL ddH2O to 1 C.C. (0500uM, 10X dilution)
- However, the precision of this dilution procedure is based on the how precise u can extract from 10,000 uM stock
- Results
|
7/30, 2008 PM 9:48
- 100, 200, 300, 400, 500 uM pi
- reaction time 10 mins
- 5N H2SO4
- Result
|
7/30, 2008 PM 10:00
- 500, 1000, 1500, 2000, 2500 uM pi (10X dilution)
- reaction time 10 mins
- 5N H2SO4
- Result
|
8/14, 2008 Trial #1
- 50, 60, 70, 80, 90, 100, 200, 300, 400, 500 uM pi
- 2 replicates
- test on Thermo Scientific Varioskan Flash
- reaction time 10 mins
|
8/14, 2008 Trial #2
- 60, 70, 80, 90, 100, 200, 300, 400, 500 uM pi
- 2 replicates
- test on Thermo Scientific Varioskan Flash
- reaction time 10 mins
|
- reaction time 20 mins
|
8/14, 2008 Trial #3 (by 佳妤)
- 50, 100, 200, 300, 400, 500, 1000, 10000 uM pi
- dilution process: from 10 mM stock to 1000 uM, 1000 uM to 500 uM, 400 uM, 300 uM, 200 uM, 100 uM and 50 uM
- test on Thermo Scientific Varioskan Flash and BECKMAN DU-640
- reaction time 10 mins
|
8/14, 2008 Trial #3 (by 志賢)
- 50, 100, 200, 300, 400, 500, 1000, 10000 uM pi
- dilution process: from 10 mM stock to 1000 uM, 500 uM, 400 uM, 300 uM, 200 uM, 100 uM and 50 uM
- test on Thermo Scientific Varioskan Flash and BECKMAN DU-640
- reaction time 10 mins
|
8/14, 2008 Trial #3 (by Jesse)
- 50, 100, 200, 300, 400, 500, 1000, 10000 uM pi
- dilution process: from 10 mM stock to 1000 uM, 500 uM, 400 uM, 300 uM, 200 uM, 100 uM and 50 uM
- test on Thermo Scientific Varioskan Flash and BECKMAN DU-640
- reaction time 10 mins
|
8/14, 2008 Trial #3 Remark
- all trials show when phosphate above 300 uM to 1000 uM, the measures are not linear as expectation
- the reaction time 10 mins may not be the best choice when we don't know anything about the kinetics
- two instruments (Thermo Scientific Varioskan Flash and BECKMAN DU-640) do not have significant measuring difference
8/21, 2008
- 500, 1000, 10000 uM pi
- dilution process: from 10 mM stock to 1000 uM, 1000 uM to 500 uM
- test on BECKMAN DU-640 to perform full spectrum scanning: 200 nm is the responsive spectrum
|
8/21, 2008
- 100, 200, 300, 400, 500, 1000, 2000, 3000, 10000 uM pi
- dilution process: from 10 mM stock to 3000, 2000, 1000 uM; 1000 uM to 500, 400, 300, 200, 100 uM
- test on BECKMAN DU-640 to perform full spectrum scanning: 200 nm is the responsive spectrum
|
- remark: This method fails in MOPS medium
8/27, 2008
- calibrated by Thermo Scientific Varioskan Flash in ddH2O
|
|
Standard Curve in MOPS
9/3, 2008
- calibrated by Thermo Scientific Varioskan Flash in MOPS medium