Team:Brown/Project/Testing

From 2008.igem.org

(Difference between revisions)
(Design V3)
Line 50: Line 50:
==Design V3==
==Design V3==
-
Having calculated the amount of E. coli bacteria cells expected in each sample and the corresponding net ionic content, we found that these values could not be accurately measured with a standard voltmeter.
+
Having calculated the amount of E. coli bacteria cells expected in each sample and the corresponding net ionic content, we found that these values could not be accurately measured with a standard voltmeter. In addition, the  electrodes of the first apparatus did not remain stationary and were difficult to manage.  They were made of a Copper alloy and were multi-stranded.  The copper was most likely performing redox chemistry.  The DNA and other charged particles released from the cells collected on the wires throughout the experiment.
-
* In addition, the  electrodes of the first apparatus did not remain stationary and were difficult to manage.  They were made of Copper alloy and multi-stranded.  The copper was most likely performing redox chemistry.  The DNA and other charged particles released from the cells collected on the wires throughout the experiment.
+
-
* The new apparatus and computer software permitted us to control the Voltage Input and Current type.  Electroporation  was always considered and a low voltage had to be used to prevent this potential problem. 
+
-
The new apparatus implemented an Alternating Current instead of a Direct Current.  The Alternating Current prevents the migration of charged molecules to one electrode, thus drastically affecting the resistance reading.       
+
-
* Our new apparatus utilized a Data Aquistion Card and the LabView Program. Team Toxipop created an electrical circuit within LabView. 
+
We designed a new system featuring a Data Acquisition Card, circuit board, amplifiers, and a LabView computer program to collect the data. The new apparatus and computer software permitted us to control the voltage being output into solution as well as the type of current (AC or DC). We elected to output an alternating current to keep the charged particles from flowing to one electrode. We also elected to output a very low voltage in order to keep the cells from electroporating.
-
* The new apparatus also consisted of a six-well plate, PDMS Mold, and Platinum Wires.         
+
 
 +
''Significant Changes:''Used a six well plate and PDMS (plastic) mold and embedded platinum wires. By using an inert metal we were able to avoid redox chemistry and by using a cast mold, we were able to keep the distances between the electrodes fixed.
 +
        
[[Image:Brown daq and cb.jpg|right|thumb|400px|New Apparatus]]
[[Image:Brown daq and cb.jpg|right|thumb|400px|New Apparatus]]
[[Image:Brown daq and mold.jpg|right|thumb|400px|New Apparatus]]
[[Image:Brown daq and mold.jpg|right|thumb|400px|New Apparatus]]
-
 
-
 
==The FINAL Apparatus==
==The FINAL Apparatus==

Revision as of 05:16, 29 October 2008




Contents

Resistance

Resistance

Team Toxipop started off the summer focusing on measuring changes in resistance. Much of our summer work was devoted to the construction of an electrical circuit and measuring apparatus unique to our purpose. Many additions were made to the apparatus each week. Below, you will find information about the different "versions" of the apparatus including final information about our switch to Conductivity.











Design V1

Design V1

Our first apparatus design was rather simple. We started with a voltmeter and a Gorillapod. We believed the Voltmeter would allow us to measure changes in resistance over time as our cells lysed. The probes of the voltmeter were attached to the Gorillapod, thus fixing the distance between them. The cell solution was contained in a Petri dish into which the probes were lowered.

Problems:The petri dish setup only allowed us to do one test at a time. Also, having to keep the entire dish open for hours on end allowed for evaporation of the solution, thus changing the volume and affecting the resistance measurements. It was also difficult to ensure that the probes were exactly the same distance apart between tests.






Design V2

Design V2

We moved to a second apparatus that fixed many of the problems we saw with Design V1. We added room for many wells, to allow for multiple tests in parallel. We also placed copper wire in each well to try and keep the electrode distance constant. Our second apparatus was made with Legos and poster board. At this point we only had a standard voltmeter that could detect large resistance changes between different salt solutions. The changes we expected to see, however, were on a much smaller scale, as determined by our mathematical calculations.

Problems: The copper wire was not stiff and would not stay at a fixed distance. We believed the voltmeter was not sensitive enough to detect magnitude of changes we were looking for. A major problem turned out to be that the voltmeter was outputting a DC (direct current) signal, which combined with copper electrodes, allowed for Redox chemistry. This led to very inconsistent readings.


Design V3

Having calculated the amount of E. coli bacteria cells expected in each sample and the corresponding net ionic content, we found that these values could not be accurately measured with a standard voltmeter. In addition, the electrodes of the first apparatus did not remain stationary and were difficult to manage. They were made of a Copper alloy and were multi-stranded. The copper was most likely performing redox chemistry. The DNA and other charged particles released from the cells collected on the wires throughout the experiment.

We designed a new system featuring a Data Acquisition Card, circuit board, amplifiers, and a LabView computer program to collect the data. The new apparatus and computer software permitted us to control the voltage being output into solution as well as the type of current (AC or DC). We elected to output an alternating current to keep the charged particles from flowing to one electrode. We also elected to output a very low voltage in order to keep the cells from electroporating.

Significant Changes:Used a six well plate and PDMS (plastic) mold and embedded platinum wires. By using an inert metal we were able to avoid redox chemistry and by using a cast mold, we were able to keep the distances between the electrodes fixed.

New Apparatus
New Apparatus

The FINAL Apparatus

  • After much work with our resistance apparatus, Team Toxipop made the transition to conductivity measurements with a Vernier Go!Link Conductivity Probe.

You can take readings in units of conductivity (µS/cm) or concentration (mg/L TDS as NaCl). The Conductivity Probe can monitor conductivity at three different sensitivity settings:

  • 0-100 mg/L TDS or 0-200 µS/cm
  • 0-1000 mg/L TDS or 0-2000 µS/cm
  • 0-10000 mg/L TDS or 0-20,000 µS/cm

Conductivity is directly proportional to concentration over the entire range.

The probe has a fast response time, reaching 98% of full value in less than 5 seconds. You can load saved calibrations using any of the Vernier data-collection programs. Alternatively, you can do a quick two-point calibration: the probe is left out of solution for one calibration point (0 µS/cm) and is placed in a known standard (1000 µS/cm solution is provided) for a second calibration point. More precise calibration in two different standards is also possible using our software.

The Conductivity Probe uses alternating current at its electrodes; this prevents polarization and electrolysis, so that solutions being tested are not fouled. Corrosion of metal electrodes is not a problem with this epoxy-body graphite electrode. It has built-in temperature compensation, which allows you to do your calibrations in the lab, and then make measurements outdoors without temperature changes affecting conductivity readings. -http://www.vernier.com/probes/con-bta.html

  • The Conductivity Probe removed any problems having do with stationary electrodes and protein and DNA buildup.